Skip to main content
TDEC
HDAQ-16250E/F module: flat dark-gray metal case with four gold SMA connectors labeled CH1–CH4

HDAQ-16250E/F

  • 8 (E) / 4 (F) synchronously sampled channels, 16-bit ADC
  • 250 kSps/CH sample rate
  • -B standard version: voltage and ICP sensor measurement
  • -S version: B/C/E/J/K/N/R/S/T thermocouple and strain measurement, differential input
  • -S version provides 18 V@50 mA/CH voltage excitation and 12 V / 6 V / 4 V / 2 V bridge excitation
  • Strain measurement pairs with the iSG-101 preamplifier or an external strain sensor
  • Channel isolation better than 90 dB, dynamic range ≥84 dB

Need selection advice or an on-site demo? Use the buttons above and our engineers will reach out shortly.

Specifications

ItemHDAQ-16250E-BHDAQ-16250F-BHDAQ-16250F-S
Measurement Performance
Channels844
Range±10 V、±5 V、±2 V、±1 V±10 V、±5 V、±2 V、±1 V±10 V、±5 V、±1 V、±0.1 V
Max Sample Rate250 KSps
Sample Rate Settings6
ADC Resolution16 bit
Signal Conditioning
ThermocoupleNoNoYes
Differential InputNoNoYes
StrainNoNoYes
Bridge Excitation Settings——12 V、6 V、4 V、2 V
Bridge Type——Determined by the strain sensor or iSG-101
ICPYes
CurrentNo
ResistanceNo
ChargeNo
AmplifierNo
Mechanical, Power & Environment
Analog Channel ConnectorSMASMA4-pin LEMO
Ingress ProtectionIP64
Dimensions120 mm×120 mm×20 mm
Operating Temperature-20 °C–60 °C
Additional Specifications
CMRR>100 dB(10 kHz)>100 dB(10 kHz)>100 dB(1 kHz)
Thermocouple Types——B, C, E, J, K, N, R, S, T
Voltage Excitation——18 V@50 mA/CH
Strain Input——External strain sensor or iSG-101 preamplifier
ICP Excitation24 V@4 mA
THD≤-83 dB
Dynamic Range≥84 dB
Channel Isolation>90 dB
Input Impedance1 MΩ
Analog Performance
Accuracy (FS)≤±0.5%FS (DC)
BandwidthDC–100kHz
SNR≥82 dB
Intra-module Channel Phase Difference0.3°@1K
Inter-module Channel Synchronisation4000nS
Filtering
Low-pass Filter10–100 kHz, 15 settings, 8th-order continuous-time elliptic
Capture & Storage
Shared Capture Buffer512 MB shared by all device channels
Storage Capacity256 GB / 1 TB
long-duration real-time recording
Internal TriggerEdge / window trigger
Trigger & Communication Interfaces
Standard External TriggerYes
Break-screen External TriggerYes
Wi-FiNo
4GNo
BeiDouNo
USBNo
EthernetNo

Accessories

Pair this device with the accessories below — click any accessory to view its specifications and detail.

Related scenarios

Outdoor Research Trials

High-precision capture of transient signals under extreme loads

Power Transient Signal Measurement

Microsecond-to-nanosecond waveform capture for insulation diagnostics and power quality

Downloads

FAQ

How do I choose between the -B and -S versions of HDAQ-16250E/F?
The -B version (E/F) measures voltage and ICP signals; the -S version (F, 4 channels) takes differential inputs for thermocouple and strain signals. Both sample at 250 KSps/CH; for 2 MSps thermocouple or strain acquisition choose HDAQ-1602F-S.