iSD-400 Distributed-Storage Dynamic Signal Analyzer
- Transient signal measurement for explosion, shock and heavy-load events
- Multi-point wireless distributed-storage synchronous measurement
- Built-in conditioning unit for direct connection of pressure, thermocouple, acceleration and other sensors
- Built-in 8 GB storage, Wi-Fi, BeiDou and rechargeable lithium battery
- iSD-401 / iSD-402 provide 1 / 2 channels respectively at up to 4 MSps/ch; iSD-404 provides 4 channels at up to 2 MSps/ch
- Ranges: ±10 V, ±2 V, ±0.5 V, ±0.1 V; resolution: 16-bit; bandwidth: 650 kHz
- Intelligent computation of characteristic parameters per relevant GJB standards
- Battery life: 12 hours for iSD-401 / iSD-402; 10 hours for iSD-404
- Applications: target vulnerability trials for explosion and shock and their dynamic signal testing; dynamic signal testing of pressure, acceleration, temperature, strain and noise
Need selection advice or an on-site demo? Use the buttons above and our engineers will reach out shortly.
Specifications
| Item | iSD-402 | iSD-404 | iSD-401 |
|---|---|---|---|
| Measurement Performance | |||
| Channels | 2 | 4 | 1 |
| Max Sample Rate | 4 MSps | 2 MSps | 4 MSps |
| Sample Rate Settings | 5 | 4 | 5 |
| ADC Resolution | 16 bit | ||
| Range | ±10 V、±2 V、±0.5V、±0.1V | ||
| Analog Performance | |||
| Inter-module Channel Synchronisation | 250nS | 500nS | 250nS |
| Accuracy (FS) | Better than 0.5%FS (DC) | ||
| Bandwidth | DC–650 kHz | ||
| SNR | Better than 80 dB | ||
| Intra-module Channel Phase Difference | 0.3°@1K | ||
| Power | |||
| Battery Life | 12 hours | 10 hours | 12 hours |
| Power Supply | External DC 24 V (PoE, 8 W) | ||
| Signal Conditioning | |||
| ICP | Yes | ||
| Thermocouple | Yes | ||
| Differential Input | No | ||
| Filtering | |||
| Low-pass Filter | 10–150 kHz, 15 settings, 8th-order continuous-time elliptic | ||
| Capture & Storage | |||
| Capture Duration per Channel | >3 s at the maximum sample rate | ||
| Storage Capacity | 8 GB (expandable to 64 GB) | ||
| Internal Trigger | Edge / window trigger | ||
| Trigger & Communication Interfaces | |||
| Standard External Trigger | Yes | ||
| Break-screen External Trigger | No | ||
| Wi-Fi | Yes | ||
| 4G | No | ||
| BeiDou | Yes | ||
| USB | No | ||
| Ethernet | Yes | ||
| Mechanical, Power & Environment | |||
| Analog Channel Connector | TRB | ||
| Ingress Protection | IP65 | ||
| Dimensions | 120 mm×80 mm×55 mm | ||
| Weight | 780g | ||
| Operating Temperature | -20 °C–60 °C | ||
| Additional Specifications | |||
| Input Impedance | 1 MΩ | ||
| Dynamic Range | Better than 90 dB | ||
| Channel Crosstalk | Better than -90 dB | ||
| Strain Measurement | With optional iSG-101 | ||
| External Trigger Level | 5 V TTL/CMOS | ||
| External Trigger Pulse Width | Greater than 100 ns | ||
| Inter-unit Sync Accuracy | 1 μs | ||
Accessories
Pair this device with the accessories below — click any accessory to view its specifications and detail.
Related scenarios
Downloads
FAQ
- How do I choose between iSD-400, the HDAQ series and iSD-101?
- iSD-401 / iSD-402 / iSD-404 provide 1 / 2 / 4 channels per unit respectively and cover large areas through distributed deployment. Choose HDAQ for centralized multi-channel measurement or iSD-101 for integrated sensing and acquisition at a single point.