Skip to main content
TDEC
HDAQ-1602E/F module: flat dark-gray metal case with eight gold SMA connectors on the side and a stacking connector on top

HDAQ-1602E/F

  • E-B: 8 channels; F-B/F-S: 4 channels; 2 MSps/ch, 16-bit
  • -B for voltage / ICP; F-S supports differential, thermocouple and strain inputs
  • ICP excitation: 24 V@4 mA
  • DC–400 kHz bandwidth; 10–150 kHz low-pass filter, 15 settings
  • Edge, window, standard external and break-screen external triggering

Need selection advice or an on-site demo? Use the buttons above and our engineers will reach out shortly.

Specifications

ItemHDAQ-1602E-BHDAQ-1602F-BHDAQ-1602F-S
Measurement Performance
Channels844
Range±10 V、±5 V、±2 V、±1 V±10 V、±5 V、±2 V、±1 V±10 V、±5 V、±1 V、±0.1 V
Max Sample Rate2 MSps
Sample Rate Settings6
ADC Resolution16 bit
Analog Performance
SNR≥82 dB≥82 dB≥82dB
Accuracy (FS)≤±0.5%FS (DC)
BandwidthDC–400 kHz
Intra-module Channel Phase Difference0.3°@1K
Inter-module Channel Synchronisation500nS
Signal Conditioning
ThermocoupleNoNoYes
Differential InputNoNoYes
StrainNoNoYes
Bridge Excitation Settings——12 V、6 V、4 V、2 V
Bridge Type——Determined by the strain sensor or iSG-101
ICPYes
CurrentNo
ResistanceNo
ChargeNo
AmplifierNo
Mechanical, Power & Environment
Analog Channel ConnectorSMASMA4-pin LEMO
Ingress ProtectionIP64
Dimensions120 mm×120 mm×20 mm
Operating Temperature-20 °C–60 °C
Additional Specifications
THD≤-110 dB≤-110 dB—
Dynamic Range≥90 dB≥90 dB—
Channel Isolation>90 dB>90 dB—
CMRR>100 dB(10 kHz)>100 dB(10 kHz)—
Input Impedance1 MΩ1 MΩ—
Strain Input——External strain sensor or iSG-101 preamplifier
ICP Excitation24 V@4 mA
Filtering
Low-pass Filter10–150 kHz, 15 settings, 8th-order continuous-time elliptic
Capture & Storage
Shared Capture Buffer512 MB shared by all device channels
Storage Capacity256 GB / 1 TB
long-duration real-time recording
Internal TriggerEdge / window trigger
Trigger & Communication Interfaces
Standard External TriggerYes
Break-screen External TriggerYes
Wi-FiNo
4GNo
BeiDouNo
USBNo
EthernetNo

Related scenarios

Outdoor Research Trials

High-precision capture of transient signals under extreme loads

Power Transient Signal Measurement

Microsecond-to-nanosecond waveform capture for insulation diagnostics and power quality

Downloads

FAQ

How do I choose between HDAQ-1602 -B and F-S?
-B accepts voltage and ICP signals. F-S provides 4 differential channels for thermocouple and strain signals and can work with an iSG-101 preamplifier. Both sample at up to 2 MSps/ch and require an HDAQ-8106 controller.